ISSN 1006-3021 CN11-3474/P
Published bimonthly started in 1979
MC-ICP-MS高精度测定Pb同位素比值
  
关键词:MC-ICP-MS  lead isotopes  correction of mass fractionation
基金项目:
作者单位E-mail
谢烈文 中国科学院地质与地球物理研究所北京100029 xieliewen@mail. igcas. ac. cn 
杨进辉 中国科学院地质与地球物理研究所北京100029  
杨岳衡 中国科学院地质与地球物理研究所北京100029  
吴福元 中国科学院地质与地球物理研究所北京100029  
摘要点击次数: 5220
全文下载次数: 2899
摘要:
High-Precision Analysis of Pb Isotope Ratios Using MC-ICP-MS
      The method for isotopic high-precision analysis of Pb by multi-collector inductively coupled plasma mass spectrometry is reported in this paper. Mass fractionation was corrected by thallium as an internal isotopic standard. The results show that the fractionation coefficients (f) of Pb and Tl are different and the f Pb/f T1 ratios are constant with changing operating parameters such as nebulizer gas, extraction potential and sampling depth. However, the is not equal to with =0.9995. The 205Tl/203Tl ratio obtained is 2.388095 by regression. The data of NIST SRM981 Pb are 208Pb/206Pb=2.16704±2, 207Pb/206Pb=0.914730±5, 206Pb/204Pb=16.9381±2, 207Pb/204Pb=15.4938±2 and 208Pb/204Pb=36.7056±7, which are consistent with the reference values measured by TIMS. The 208Pb/206Pb ratio, however, is a bit lower. The precision of the analyses made by the authors seem to be 4~11 factors better than those from TIMS by Baker et al. (2004).
ME Liewen,YANG Jinhui,YANG Yueheng,WU Fuyuan.2005.High-Precision Analysis of Pb Isotope Ratios Using MC-ICP-MS[J].Acta Geoscientica Sinica,26(z1):23.
查看全文  查看/发表评论  下载PDF阅读器
Copyright©2008 All Rights Reserved
Sponsored by: